21 CFR Part 820 · Quality System Regulation (before Feb 2, 2026)

21 CFR 820.72

Inspection, measuring, and test equipment

21 CFR 820.72 is the regulation titled “Inspection, measuring, and test equipment” in 21 CFR Part 820, Quality System Regulation (before Feb 2, 2026). Since 2021, 10 FDA warning letters have cited it.

Part 820 changed when FDA's Quality Management System Regulation took effect on February 2, 2026. This is the earlier text these letters cite.

Letters citing it
10 since 2021
Jan 1 to Oct 30, 2026
0 against 2 on the same dates in 2025
Most often cited with
21 CFR 820.30, in 9 letters
Closed out
30% of these letters have a published close-out
See the 10 letters citing 21 CFR 820.72 →

Letters citing it, by year issued

All are medical device letters. 2026 runs to Oct 30, 2026. Pick a year to see its letters.

  1. 02021
  2. 22022
  3. 52023
  4. 12024
  5. 22025
  6. 02026

What FDA writes when it cites it

The finding headings that cite 21 CFR 820.72 most often, as written in the letters.

  • 2 lettersFailure to ensure that all inspection, measuring, and test equipment, including mechanical, automated, or electronic inspection and test equipment, is suitable for its intended purposes and is capable of producing valid…
  • 1 letterFailure to adequately document equipment inspections, checks and maintenance activities, as required by 21 CFR 820.72(a)
  • 1 letterFailure to adequately establish and maintain procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained as required by 21 CFR 820.72(a)

The regulation

Read the text of 21 CFR 820.72 (eCFR, as of Jun 1, 2025)

(a) Control of inspection, measuring, and test equipment. Each manufacturer shall ensure that all inspection, measuring, and test equipment, including mechanical, automated, or electronic inspection and test equipment, is suitable for its intended purposes and is capable of producing valid results. Each manufacturer shall establish and maintain procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained. The procedures shall include provisions for handling, preservation, and storage of equipment, so that its accuracy and fitness for use are maintained. These activities shall be documented.

(b) Calibration. Calibration procedures shall include specific directions and limits for accuracy and precision. When accuracy and precision limits are not met, there shall be provisions for remedial action to reestablish the limits and to evaluate whether there was any adverse effect on the device's quality. These activities shall be documented.

(1) Calibration standards. Calibration standards used for inspection, measuring, and test equipment shall be traceable to national or international standards. If national or international standards are not practical or available, the manufacturer shall use an independent reproducible standard. If no applicable standard exists, the manufacturer shall establish and maintain an in-house standard.

(2) Calibration records. The equipment identification, calibration dates, the individual performing each calibration, and the next calibration date shall be documented. These records shall be displayed on or near each piece of equipment or shall be readily available to the personnel using such equipment and to the individuals responsible for calibrating the equipment.

Text from eCFR as of Jun 1, 2025.

Cited alongside it

Regulations that appear in the same letters as 21 CFR 820.72.

Latest letters citing it

The 3 letters citing 21 CFR 820.72 that FDA posted most recently.

See the 10 letters citing 21 CFR 820.72 →
For regulated teams

Put this kind of reading to
work on your own records.

Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.

Counts include every citation in a letter, as the Letters filter does. How this database is built.

Every letter by dateHow this database is built

FDA Warning Letter Alerts

Share this view

Send the same filters, search and view to a colleague.

You choose what to post or send. Sharing a link does not change access to this site.