21 CFR Part 820 · Quality System Regulation (before Feb 2, 2026)

21 CFR 820.70

Production and process controls

21 CFR 820.70 is the regulation titled “Production and process controls” in 21 CFR Part 820, Quality System Regulation (before Feb 2, 2026). Since 2021, 31 FDA warning letters have cited it.

Part 820 changed when FDA's Quality Management System Regulation took effect on February 2, 2026. This is the earlier text these letters cite.

Letters citing it
31 since 2021
Jan 1 to Oct 30, 2026
4 against 7 on the same dates in 2025
Most often cited with
21 CFR 820.30, in 20 letters
Closed out
32% of these letters have a published close-out
See the 31 letters citing 21 CFR 820.70 →

Letters citing it, by year issued

All are medical device letters. 2026 runs to Oct 30, 2026. Pick a year to see its letters.

  1. 52021
  2. 22022
  3. 72023
  4. 42024
  5. 92025
  6. 42026

What FDA writes when it cites it

The finding headings that cite 21 CFR 820.70 most often, as written in the letters.

  • 3 lettersFailure to develop, conduct, control, and monitor production processes to ensure that a device conforms to its specifications, as required by 21 CFR 820.70(a)
  • 2 lettersFailure to establish and maintain procedures to adequately control environmental conditions, which could reasonably be expected to have an adverse effect on product quality and periodically inspect the environmental…
  • 2 lettersFailure to validate software used as part of the quality system for its intended use according to an established protocol, as required by 21 CFR 820.70(i)

The regulation

Read the text of 21 CFR 820.70 (eCFR, as of Jun 1, 2025)

(a) General. Each manufacturer shall develop, conduct, control, and monitor production processes to ensure that a device conforms to its specifications. Where deviations from device specifications could occur as a result of the manufacturing process, the manufacturer shall establish and maintain process control procedures that describe any process controls necessary to ensure conformance to specifications. Where process controls are needed they shall include:

(1) Documented instructions, standard operating procedures (SOP's), and methods that define and control the manner of production;

(2) Monitoring and control of process parameters and component and device characteristics during production;

(3) Compliance with specified reference standards or codes;

(4) The approval of processes and process equipment; and

(5) Criteria for workmanship which shall be expressed in documented standards or by means of identified and approved representative samples.

(b) Production and process changes. Each manufacturer shall establish and maintain procedures for changes to a specification, method, process, or procedure. Such changes shall be verified or where appropriate validated according to § 820.75, before implementation and these activities shall be documented. Changes shall be approved in accordance with § 820.40.

(c) Environmental control. Where environmental conditions could reasonably be expected to have an adverse effect on product quality, the manufacturer shall establish and maintain procedures to adequately control these environmental conditions. Environmental control system(s) shall be periodically inspected to verify that the system, including necessary equipment, is adequate and functioning properly. These activities shall be documented and reviewed.

(d) Personnel. Each manufacturer shall establish and maintain requirements for the health, cleanliness, personal practices, and clothing of personnel if contact between such personnel and product or environment could reasonably be expected to have an adverse effect on product quality. The manufacturer shall ensure that maintenance and other personnel who are required to work temporarily under special environmental conditions are appropriately trained or supervised by a trained individual.

(e) Contamination control. Each manufacturer shall establish and maintain procedures to prevent contamination of equipment or product by substances that could reasonably be expected to have an adverse effect on product quality.

(f) Buildings. Buildings shall be of suitable design and contain sufficient space to perform necessary operations, prevent mixups, and assure orderly handling.

(g) Equipment. Each manufacturer shall ensure that all equipment used in the manufacturing process meets specified requirements and is appropriately designed, constructed, placed, and installed to facilitate maintenance, adjustment, cleaning, and use.

(1) Maintenance schedule. Each manufacturer shall establish and maintain schedules for the adjustment, cleaning, and other maintenance of equipment to ensure that manufacturing specifications are met. Maintenance activities, including the date and individual(s) performing the maintenance activities, shall be documented.

(2) Inspection. Each manufacturer shall conduct periodic inspections in accordance with established procedures to ensure adherence to applicable equipment maintenance schedules. The inspections, including the date and individual(s) conducting the inspections, shall be documented.

(3) Adjustment. Each manufacturer shall ensure that any inherent limitations or allowable tolerances are visibly posted on or near equipment requiring periodic adjustments or are readily available to personnel performing these adjustments.

(h) Manufacturing material. Where a manufacturing material could reasonably be expected to have an adverse effect on product quality, the manufacturer shall establish and maintain procedures for the use and removal of such manufacturing material to ensure that it is removed or limited to an amount that does not adversely affect the device's quality. The removal or reduction of such manufacturing material shall be documented.

(i) Automated processes. When computers or automated data processing systems are used as part of production or the quality system, the manufacturer shall validate computer software for its intended use according to an established protocol. All software changes shall be validated before approval and issuance. These validation activities and results shall be documented.

Text from eCFR as of Jun 1, 2025.

Cited alongside it

Regulations that appear in the same letters as 21 CFR 820.70.

Latest letters citing it

The 3 letters citing 21 CFR 820.70 that FDA posted most recently.

See the 31 letters citing 21 CFR 820.70 →
For regulated teams

Put this kind of reading to
work on your own records.

Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.

Counts include every citation in a letter, as the Letters filter does. How this database is built.

Every letter by dateHow this database is built