21 CFR 812.20
Application
21 CFR 812.20 is the regulation titled “Application” in 21 CFR Part 812, Investigational Device Exemptions. Since 2021, 1 FDA warning letter has cited it.
- Letters citing it
- 1 since 2021
- Jan 1 to Oct 30, 2026
- 0 against 1 on the same dates in 2025
- Most often cited with
- 21 CFR 812.100, in 1 letter
- Closed out
- 0% of these letters have a published close-out
Letters citing it, by year issued
It is a medical device letter. 2026 runs to Oct 30, 2026. Pick a year to see its letters.
- 02021
- 02022
- 02023
- 02024
- 12025
- 02026
What FDA writes when it cites it
The finding headings that cite 21 CFR 812.20 most often, as written in the letters.
- 1 letterFailure to submit an IDE application to FDA and failure to obtain FDA approval prior to beginning an investigation for which FDA’s approval is required [21 CFR 812.20(a)(1) and (2)
The regulation
Read the text of 21 CFR 812.20 (eCFR, as of Sep 1, 2026)
(a) Submission. (1) A sponsor shall submit an application to FDA if the sponsor intends to use a significant risk device in an investigation, intends to conduct an investigation that involves an exception from informed consent under § 50.24 of this chapter, or if FDA notifies the sponsor that an application is required for an investigation.
(2) A sponsor shall not begin an investigation for which FDA's approval of an application is required until FDA has approved the application.
(3) A sponsor shall submit a signed “Application for an Investigational Device Exemption” (IDE application), together with accompanying materials in electronic format, to one of the addresses in § 812.19, and if eCopy by registered mail or by hand. Subsequent correspondence concerning an application or a supplemental application shall be submitted in electronic format and if eCopy by registered mail or by hand.
(4)(i) A sponsor shall submit a separate IDE for any clinical investigation involving an exception from informed consent under § 50.24 of this chapter. Such a clinical investigation is not permitted to proceed without the prior written authorization of FDA. FDA shall provide a written determination 30 days after FDA receives the IDE or earlier.
(ii) If the investigation involves an exception from informed consent under § 50.24 of this chapter, the sponsor shall prominently identify on the cover sheet that the investigation is subject to the requirements in § 50.24 of this chapter.
(b) Contents. An IDE application shall include, in the following order:
(1) The name and address of the sponsor.
(2) A complete report of prior investigations of the device and an accurate summary of those sections of the investigational plan described in § 812.25(a) through (e) or, in lieu of the summary, the complete plan. The sponsor shall submit to FDA a complete investigational plan and a complete report of prior investigations of the device if no IRB has reviewed them, if FDA has found an IRB's review inadequate, or if FDA requests them.
(3) A description of the methods, facilities, and controls used for the manufacture, processing, packing, storage, and, where appropriate, installation of the device, in sufficient detail so that a person generally familiar with good manufacturing practices can make a knowledgeable judgment about the quality control used in the manufacture of the device.
(4) An example of the agreements to be entered into by all investigators to comply with investigator obligations under this part, and a list of the names and addresses of all investigators who have signed the agreement.
(5) A certification that all investigators who will participate in the investigation have signed the agreement, that the list of investigators includes all the investigators participating in the investigation, and that no investigators will be added to the investigation until they have signed the agreement.
(6) A list of the name, address, and chairperson of each IRB that has been or will be asked to review the investigation and a certification of the action concerning the investigation taken by each such IRB.
(7) The name and address of any institution at which a part of the investigation may be conducted that has not been identified in accordance with paragraph (b)(6) of this section.
(8) If the device is to be sold, the amount to be charged and an explanation of why sale does not constitute commercialization of the device.
(9) A claim for categorical exclusion under § 25.30 or § 25.34 or an environmental assessment under § 25.40.
(10) Copies of all labeling for the device.
(11) Copies of all forms and informational materials to be provided to subjects to obtain informed consent.
(12) Any other relevant information FDA requests for review of the application.
(c) Additional information. FDA may request additional information concerning an investigation or revision in the investigational plan. The sponsor may treat such a request as a disapproval of the application for purposes of requesting a hearing under part 16.
(d) Information previously submitted. Information previously submitted to the Center for Devices and Radiological Health, the Center for Biologics Evaluation and Research, or the Center for Drug Evaluation and Research, as applicable, in accordance with this chapter ordinarily need not be resubmitted, but may be incorporated by reference.
[45 FR 3751, Jan. 18, 1980, as amended at 46 FR 8956, Jan. 27, 1981; 50 FR 16669, Apr. 26, 1985; 53 FR 11252, Apr. 6, 1988; 61 FR 51530, Oct. 2, 1996; 62 FR 40600, July 29, 1997; 64 FR 10942, Mar. 8, 1999; 73 FR 49942, Aug. 25, 2008; 84 FR 68339, Dec. 16, 2019]
Text from eCFR as of Sep 1, 2026.
Cited alongside it
Regulations that appear in the same letters as 21 CFR 812.20.
- 21 CFR 812.100General responsibilities of investigators1 letter cites both
- 21 CFR 812.140Records1 letter cites both
- 21 CFR 812.150Reports1 letter cites both
- 21 CFR 812.3Definitions1 letter cites both
- 21 CFR 812.40General responsibilities of sponsors1 letter cites both
- 21 CFR 812.46Monitoring investigations1 letter cites both
- 21 CFR 878.4490Absorbable hemostatic agent and dressing1 letter cites both
Latest letters citing it
The letter citing 21 CFR 812.20 that FDA posted most recently.
See the letter citing 21 CFR 812.20 →Put this kind of reading to
work on your own records.
Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.
Counts include every citation in a letter, as the Letters filter does. How this database is built.