21 CFR 812.150
Reports
21 CFR 812.150 is the regulation titled “Reports” in 21 CFR Part 812, Investigational Device Exemptions. Since 2021, 1 FDA warning letter has cited it.
- Letters citing it
- 1 since 2021
- Jan 1 to Oct 30, 2026
- 0 against 1 on the same dates in 2025
- Most often cited with
- 21 CFR 812.100, in 1 letter
- Closed out
- 0% of these letters have a published close-out
Letters citing it, by year issued
It is a medical device letter. 2026 runs to Oct 30, 2026. Pick a year to see its letters.
- 02021
- 02022
- 02023
- 02024
- 12025
- 02026
What FDA writes when it cites it
The finding headings that cite 21 CFR 812.150 most often, as written in the letters.
- 1 letterFailure to immediately conduct an evaluation of any unanticipated adverse device effects and failure to report results of such an evaluation to FDA and to all reviewing IRB’s and participating investigators within 10…
The regulation
Read the text of 21 CFR 812.150 (eCFR, as of Sep 1, 2026)
(a) Investigator reports. An investigator shall prepare and submit the following complete, accurate, and timely reports:
(1) Unanticipated adverse device effects. An investigator shall submit to the sponsor and to the reviewing IRB a report of any unanticipated adverse device effect occurring during an investigation as soon as possible, but in no event later than 10 working days after the investigator first learns of the effect.
(2) Withdrawal of IRB approval. An investigator shall report to the sponsor, within 5 working days, a withdrawal of approval by the reviewing IRB of the investigator's part of an investigation.
(3) Progress. An investigator shall submit progress reports on the investigation to the sponsor, the monitor, and the reviewing IRB at regular intervals, but in no event less often than yearly.
(4) Deviations from the investigational plan. An investigator shall notify the sponsor and the reviewing IRB (see § 56.108(a) (3) and (4)) of any deviation from the investigational plan to protect the life or physical well-being of a subject in an emergency. Such notice shall be given as soon as possible, but in no event later than 5 working days after the emergency occurred. Except in such an emergency, prior approval by the sponsor is required for changes in or deviations from a plan, and if these changes or deviations may affect the scientific soundness of the plan or the rights, safety, or welfare of human subjects, FDA and IRB in accordance with § 812.35(a) also is required.
(5) Informed consent. If an investigator uses a device without obtaining informed consent, the investigator shall report such use to the sponsor and the reviewing IRB within 5 working days after the use occurs.
(6) Final report. An investigator shall, within 3 months after termination or completion of the investigation or the investigator's part of the investigation, submit a final report to the sponsor and the reviewing IRB.
(7) Other. An investigator shall, upon request by a reviewing IRB or FDA, provide accurate, complete, and current information about any aspect of the investigation.
(b) Sponsor reports. A sponsor shall prepare and submit the following complete, accurate, and timely reports:
(1) Unanticipated adverse device effects. A sponsor who conducts an evaluation of an unanticipated adverse device effect under § 812.46(b) shall report the results of such evaluation to FDA and to all reviewing IRB's and participating investigators within 10 working days after the sponsor first receives notice of the effect. Thereafter the sponsor shall submit such additional reports concerning the effect as FDA requests.
(2) Withdrawal of IRB approval. A sponsor shall notify FDA and all reviewing IRB's and participating investigators of any withdrawal of approval of an investigation or a part of an investigation by a reviewing IRB within 5 working days after receipt of the withdrawal of approval.
(3) Withdrawal of FDA approval. A sponsor shall notify all reviewing IRB's and participating investigators of any withdrawal of FDA approval of the investigation, and shall do so within 5 working days after receipt of notice of the withdrawal of approval.
(4) Current investigator list. A sponsor shall submit to FDA, at 6-month intervals, a current list of the names and addresses of all investigators participating in the investigation. The sponsor shall submit the first such list 6 months after FDA approval.
(5) Progress reports. At regular intervals, and at least yearly, a sponsor shall submit progress reports to all reviewing IRB's. In the case of a significant risk device, a sponsor shall also submit progress reports to FDA. A sponsor of a treatment IDE shall submit semi-annual progress reports to all reviewing IRB's and FDA in accordance with § 812.36(f) and annual reports in accordance with this section.
(6) Recall and device disposition. A sponsor shall notify FDA and all reviewing IRB's of any request that an investigator return, repair, or otherwise dispose of any units of a device. Such notice shall occur within 30 working days after the request is made and shall state why the request was made.
(7) Final report. In the case of a significant risk device, the sponsor shall notify FDA within 30 working days of the completion or termination of the investigation and shall submit a final report to FDA and all reviewing the IRB's and participating investigators within 6 months after completion or termination. In the case of a device that is not a significant risk device, the sponsor shall submit a final report to all reviewing IRB's within 6 months after termination or completion.
(8) Informed consent. A sponsor shall submit to FDA a copy of any report by an investigator under paragraph (a)(5) of this section of use of a device without obtaining informed consent, within 5 working days of receipt of notice of such use.
(9) Significant risk device determinations. If an IRB determines that a device is a significant risk device, and the sponsor had proposed that the IRB consider the device not to be a significant risk device, the sponsor shall submit to FDA a report of the IRB's determination within 5 working days after the sponsor first learns of the IRB's determination.
(10) Other. A sponsor shall, upon request by a reviewing IRB or FDA, provide accurate, complete, and current information about any aspect of the investigation.
[45 FR 3751, Jan. 18, 1980, as amended at 45 FR 58843, Sept. 5, 1980; 48 FR 15622, Apr. 12, 1983; 62 FR 48948, Sept. 18, 1997]
Text from eCFR as of Sep 1, 2026.
Cited alongside it
Regulations that appear in the same letters as 21 CFR 812.150.
- 21 CFR 812.100General responsibilities of investigators1 letter cites both
- 21 CFR 812.140Records1 letter cites both
- 21 CFR 812.20Application1 letter cites both
- 21 CFR 812.3Definitions1 letter cites both
- 21 CFR 812.40General responsibilities of sponsors1 letter cites both
- 21 CFR 812.46Monitoring investigations1 letter cites both
- 21 CFR 878.4490Absorbable hemostatic agent and dressing1 letter cites both
Latest letters citing it
The letter citing 21 CFR 812.150 that FDA posted most recently.