21 CFR Part 1271 · Human Cells, Tissues, and Cellular and Tissue-Based Products

21 CFR 1271.85

What donor testing is required for different types of cells and tissues?

21 CFR 1271.85 is the regulation titled “What donor testing is required for different types of cells and tissues?” in 21 CFR Part 1271, Human Cells, Tissues, and Cellular and Tissue-Based Products. Since 2021, 12 FDA warning letters have cited it.

Letters citing it
12 since 2021
Jan 1 to Oct 30, 2026
3 against 2 on the same dates in 2025
Most often cited with
21 CFR 1271.60, in 12 letters
Closed out
8% of these letters have a published close-out
See the 12 letters citing 21 CFR 1271.85 →

Letters citing it, by year issued

All are biologics letters. 2026 runs to Oct 30, 2026. Pick a year to see its letters.

  1. 12021
  2. 32022
  3. 02023
  4. 22024
  5. 32025
  6. 32026

What FDA writes when it cites it

The finding headings that cite 21 CFR 1271.85 most often, as written in the letters.

  • 2 lettersFailure to test a specimen from an anonymous or directed reproductive donor of cells or tissue, whether viable or non-viable, for evidence of infection due to relevant communicable disease agents [21 CFR 1271.85(a)].…
  • 1 letterFailure to determine as ineligible a donor whose specimen tests reactive on a screening test for a communicable disease agent for which testing is required under 21 CFR 1271.85 [21 CFR 1271.80(d)(1)]. For example, a…
  • 1 letterFailure to determine whether a donor is eligible based upon the results of donor screening in accordance with 21 CFR 1271.75 and donor testing in accordance with 21 CFR 1271.80 and 21 CFR 1271.85 [21 CFR 1271.50(a)].…

The regulation

Read the text of 21 CFR 1271.85 (eCFR, as of Sep 1, 2026)

(a) All donors. To adequately and appropriately reduce the risk of transmission of relevant communicable diseases, and except as provided under § 1271.90, you must test a specimen from the donor of cells or tissue, whether viable or nonviable, for evidence of infection due to relevant communicable disease agents, including:

(1) Human immunodeficiency virus, type 1;

(2) Human immunodeficiency virus, type 2;

(3) Hepatitis B virus;

(4) Hepatitis C virus; and

(5) Treponema pallidum.

(b) Donors of viable, leukocyte-rich cells or tissue. In addition to the relevant communicable disease agents for which testing is required under paragraph (a) of this section, and except as provided under § 1271.90,

(1) You must test a specimen from the donor of viable, leukocyte-rich cells or tissue to adequately and appropriately reduce the risk of transmission of relevant cell-associated communicable diseases, including:

(i) Human T-lymphotropic virus, type I; and

(ii) Human T-lymphotropic virus, type II.

(2) You must test a specimen from the donor of viable, leukocyte-rich cells or tissue for evidence of infection due to cytomegalovirus (CMV), to adequately and appropriately reduce the risk of transmission. You must establish and maintain a standard operating procedure governing the release of an HCT/P from a donor whose specimen tests reactive for CMV.

(c) Donors of reproductive cells or tissue. In addition to the communicable disease agents for which testing is required under paragraphs (a) and (b) of this section, as applicable, and except as provided under § 1271.90, you must test a specimen from the donor of reproductive cells or tissue to adequately and appropriately reduce the risk of transmission of relevant communicable disease agents of the genitourinary tract. Such testing must include testing for the communicable disease agents listed in paragraphs (c)(1) and (c)(2) of this section. However, if the reproductive cells or tissues are recovered by a method that ensures freedom from contamination of the cells or tissue by infectious disease organisms that may be present in the genitourinary tract, then testing for the communicable disease agents listed in paragraphs (c)(1) and (c)(2) of this section is not required. Communicable disease agents of the genitourinary tract for which you must test include:

(1) Chlamydia trachomatis; and

(2) Neisseria gonorrhea.

(d) Retesting anonymous semen donors. Except as provided under § 1271.90 and except for directed reproductive donors as defined in § 1271.3(l), at least 6 months after the date of donation of semen from anonymous donors, you must collect a new specimen from the donor and test it for evidence of infection due to the communicable disease agents for which testing is required under paragraphs (a), (b), and (c) of this section.

(e) Dura mater. For donors of dura mater, you must perform an adequate assessment designed to detect evidence of transmissible spongiform encephalopathy.

Text from eCFR as of Sep 1, 2026.

Cited alongside it

Regulations that appear in the same letters as 21 CFR 1271.85.

Latest letters citing it

The 3 letters citing 21 CFR 1271.85 that FDA posted most recently.

See the 12 letters citing 21 CFR 1271.85 →
For regulated teams

Put this kind of reading to
work on your own records.

Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.

Counts include every citation in a letter, as the Letters filter does. How this database is built.

Every letter by dateHow this database is built

FDA Warning Letter Alerts

Share this view

Send the same filters, search and view to a colleague.

You choose what to post or send. Sharing a link does not change access to this site.