21 CFR Part 878 · General and Plastic Surgery Devices

21 CFR 878.4430

Microneedling device for aesthetic use

21 CFR 878.4430 is the regulation titled “Microneedling device for aesthetic use” in 21 CFR Part 878, General and Plastic Surgery Devices. Since 2021, 1 FDA warning letter has cited it.

Letters citing it
1 since 2021
Jan 1 to Oct 30, 2026
0 against 0 on the same dates in 2025
Most often cited with
21 CFR 3.2, in 1 letter
Closed out
0% of these letters have a published close-out
See the letter citing 21 CFR 878.4430 →

Letters citing it, by year issued

It is a medical device letter. 2026 runs to Oct 30, 2026. Pick a year to see its letters.

  1. 12021
  2. 02022
  3. 02023
  4. 02024
  5. 02025
  6. 02026

The regulation

Read the text of 21 CFR 878.4430 (eCFR, as of Sep 1, 2026)

(a) Identification. A microneedling device for aesthetic use is a device using one or more needles to mechanically puncture and injure skin tissue for aesthetic use. This classification does not include devices intended for transdermal delivery of topical products such as cosmetics, drugs, or biologics.

(b) Classification. Class II (special controls). The special controls for this device are:

(1) The technical specifications and needle characteristics must be identified, including needle length, geometry, maximum penetration depth, and puncture rate.

(2) Non-clinical performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:

(i) Accuracy of needle penetration depth and puncture rate;

(ii) Safety features built into the device to protect against cross-contamination, including fluid ingress protection; and

(iii) Identification of the maximum safe needle penetration depth for the device for the labeled indications for use.

(3) Performance data must demonstrate the sterility of the patient-contacting components of the device.

(4) Performance data must support the shelf life of the device by demonstrating continued sterility, package integrity, and device functionality over the intended shelf life.

(5) Performance data must demonstrate the electrical safety and electromagnetic compatibility (EMC) of all electrical components of the device.

(6) Software verification, validation, and hazard analysis must be performed for all software components of the device.

(7) The patient-contacting components of the device must be demonstrated to be biocompatible.

(8) Performance data must validate the cleaning and disinfection instructions for reusable components of the device.

(9) Labeling must include the following:

(i) Information on how to operate the device and its components and the typical course of treatment;

(ii) A summary of the device technical parameters, including needle length, needle geometry, maximum penetration depth, and puncture rate;

(iii) Validated methods and instructions for reprocessing of any reusable components;

(iv) Disposal instructions; and

(v) A shelf life.

(10) Patient labeling must be provided and must include:

(i) Information on how the device operates and the typical course of treatment;

(ii) The probable risks and benefits associated with use of the device; and

(iii) Postoperative care instructions.

[83 FR 26577, June 8, 2018]

Text from eCFR as of Sep 1, 2026.

Cited alongside it

Regulations that appear in the same letters as 21 CFR 878.4430.

Latest letters citing it

The letter citing 21 CFR 878.4430 that FDA posted most recently.

See the letter citing 21 CFR 878.4430 →
For regulated teams

Put this kind of reading to
work on your own records.

Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.

Counts include every citation in a letter, as the Letters filter does. How this database is built.

Every letter by dateHow this database is built

FDA Warning Letter Alerts

Share this view

Send the same filters, search and view to a colleague.

You choose what to post or send. Sharing a link does not change access to this site.