Devices

Sedecal S.A.

FDA issued a warning letter to Sedecal S.A. (ALGETE, Madrid, Spain) on May 5, 2025, citing 21 CFR 820.100, 21 CFR 820.30, 21 CFR 806.10, 21 CFR 1020.30 and 1 more regulation. Subject: CGMP, QSR, Medical Devices, Adulterated.

From the letter: “your firm failed to document an assessment of the risk of the affected units in the field for this issue, when investigating the cause of nonconformities relating to the product.”

ALGETE, Madrid, SpainIssued May 5, 2025Posted May 27, 2025No close-out recorded

At a glance

What a reviewer notes before reading: findings, citations, the response deadline and what the letter warns of.

Violations
4 findings · 4 issue types
Response due
15 business days after receipt of this letter
FDA asks for
7 actions
Named as possible actions
Refusal of admission
Issued by
Center for Devices and Radiological Health
Published status
No close-out recorded

AI assisted findings

Model-assisted reading of each issue. Open one to see the FDA text it comes from, and check it against the full letter.

01CAPA · 21 CFR 820.100(a)CAPA 20/028 failed to document a risk assessment for mobile x-ray units with unintended movement, despite affecting all models.
your firm failed to document an assessment of the risk of the affected units in the field for this issue, when investigating the cause of nonconformities relating to the product.
02Design controls · 21 CFR 820.30(g)Design validation risk analysis for MobileDiagnost wDR omitted the increased hazard severity (life threatening) from a recall.
the risk analysis for MobileDiagnost wDR (Document IH0532RA_Annex, dated April 26, 2022) fails to include this increased hazard severity (life threatening) to ensure that appropriate risk control measures are implemented.
03Complaints & adverse event reporting · 21 CFR 806.10Failed to submit a written report to FDA for a correction/removal initiated to reduce risk to health for unintended movement.
your firm failed to submit a report of correction or removal to FDA for these actions.
04Registration, listing & premarket · 21 CFR 1002.30(a)(1)Released wDR2.2 for distribution without completing the required performance standard checklist, which was not done until later.
the wDR2.2 was approved and released for distribution in the US without documentation to show that the required performance standard checklist (Technical Report Subchapter J of 21 CFR MobileDiagnost wDR 2.2, Doc #R01005_21CKL) was completed to demonstrate compliance with the performance standard requirements prior to release of the product. The performance standard checklist was not completed until Jan. 14, 2025.

What FDA requested

  • Revise procedures PG-031 Customer complaint and PG-020 Risk management procedures to ensure that an investigation is documented properly.
  • Revise procedure PG-019 Corrective and Preventive action and conduct training for the relevant personnel.
  • Include revised procedures in the response for evaluation.
  • Conduct a retrospective review of related documentation to ensure compliance with the updated procedures and how these will be implemented moving forward.
  • Re-evaluate and document the risk analysis for the recall.
  • Revise procedures PE-062 Design Verification, PG-044 Organization of Project Teams among others.
  • Include documentation of corrective action, retrospective review, and training.

FDA Warning Letter

22 paragraphs · 1 section

Text as published on FDA.gov, retrieved Sep 8, 2026. Layout added for reading; the wording is FDA’s. View on FDA.gov (opens in a new tab)

DatedMay 5, 2025
AddressedDear Mr. Martinez García

1Letter text

16 paragraphs

During an inspection of your firm located in Madrid, Spain on January 20, 2025, through January 24, 2025, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures DX-D 100, DX-D 600, and MobileDiagnost wDR systems. Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or function of the body.

This inspection revealed that these devices are adulterated within the meaning of section 501(h) of the Act, 21 U.S.C. § 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the current good manufacturing practice requirements of the Quality System regulation found at Title 21, Code of Federal Regulations (CFR), Part 820. We received a response from Maria Luisa Gomez de Aguero, Director of Quality and Regulatory Affairs dated February 14, 2025, concerning our investigator’s observations noted on the Form FDA 483 (FDA 483), List of Inspectional Observations, which was issued to your firm. We address this response below, in relation to each of the noted violations. These violations include, but are not limited to, the following:

1. Failure to adequately establish and maintain procedures for implementing corrective and preventative action (CAPA), as required by 21 CFR 820.100(a). Specifically, your firm initiated CAPA 20/028 to address customer complaints reporting DXD100 mobile x-ray units self-propelling crashing into x-ray control panel in one case and almost running into the patient, in the other case. This issue affects all mobile x-ray unit models. However, your firm failed to document an assessment of the risk of the affected units in the field for this issue, when investigating the cause of nonconformities relating to the product.

We reviewed your response and conclude that the adequacy of your firm’s response cannot be determined at this time. You stated that you will revise procedures PG-031 Customer complaint and PG-020 Risk management procedures to ensure that an investigation is documented properly. You also stated that you will revise procedure PG-019 Corrective and Preventive action and conduct training for the relevant personnel. Although your firm’s response states you will revise the relevant procedures, the revised procedures were not included in the response for evaluation. Your firm’s response does not include a retrospective review of related documentation to ensure compliance with the updated procedures and how these will be implemented moving forward. Documentation of corrective action, retrospective review as well as training is needed in order to adequately address this observation.

2. Failure to adequately establish and maintain procedures for validating the device design, as required by 21 CFR 820.30(g). Design validation shall include risk analysis, where appropriate. Specifically, your firm conducted a recall to address unintended movement of MobileDiagnost wDR that could result in life threatening situations and cause the device to stop working as intended. However, the risk analysis for MobileDiagnost wDR (Document IH0532RA_Annex, dated April 26, 2022) fails to include this increased hazard severity (life threatening) to ensure that appropriate risk control measures are implemented.

We reviewed your response and conclude that the adequacy of your firm’s response cannot be determined at this time. Your response states that you will re-evaluate and document the risk analysis for the recall. You also state that you will revise procedures PG-031 Customer complaint and PG-020 Risk management procedures to account for risk analysis. Although your firm’s response states you will revise the relevant procedures, the revised procedures were not included in the response for evaluation. Your firm’s response does not include a retrospective review of related documentation to ensure compliance with the updated procedures and how these will be implemented moving forward. Documentation of corrective action, retrospective review as well as training is needed in order to adequately address this observation.

Our inspection also revealed that your firm’s DX-D100 devices are misbranded under section 502(t)(2) of the Act, 21 U.S.C. § 352(t)(2), in that your firm failed or refused to furnish material or information respecting the device that is required by or under section 519 of the Act, 21 U.S.C. § 360i, and 21 CFR Part 806 – Medical Device; Reports of Corrections and Removals. Significant violations include, but are not limited to, the following:

3. Failure to submit a written report of a correction or removal of a device to FDA, initiated to reduce a risk to health or to remedy a violation which may present a risk to health, as required by 21 CFR 806.10. Specifically, your firm initiated CAPA 20/028 to address customer complaints reporting DXD100 mobile x-ray units self-propelling crashing into x-ray control panel in one case and almost running into the patient, in the other case. This issue affects all mobile x-ray unit models. Your firm implemented two corrective actions to address this issue and issued a service note to be distributed to customers. However, your firm failed to submit a report of correction or removal to FDA for these actions.

We reviewed your response to Observation 1 and conclude it is not adequate. Your investigation of the issue concluded that this issue does not pose a risk to safety due to the low speed of the movement and did not consider it as a reportable recall. However, your firm implemented two corrections to address this issue and issued a service note to be distributed to the customers. Our assessment concluded that the unintended movement presents a risk to health and your firm’s actions were initiated to reduce the risk to health.

Your firm’s devices are also electronic products subject to compliance with Subchapter C of the Act, Electronic Product Radiation Control, the requirements at 21 CFR 1000-1005, and the performance standards at 21 CFR 1020.30 and 1020.31. Your firm failed to comply with the regulations regarding reporting and recordkeeping.

4. Failure to conduct certification of an electronic product, based upon a testing program which is in accordance with good manufacturing practices, as required by 21 CFR 1002.30(a)(1). Specifically, a review of the design control records revealed that the wDR2.2 was approved and released for distribution in the US without documentation to show that the required performance standard checklist (Technical Report Subchapter J of 21 CFR MobileDiagnost wDR 2.2, Doc #R01005_21CKL) was completed to demonstrate compliance with the performance standard requirements prior to release of the product. The performance standard checklist was not completed until Jan. 14, 2025.

We reviewed your response and conclude that the adequacy of your firm’s response cannot be determined at this time. You stated that you will review and revise procedures PE-062 Design Verification, PG-044 Organization of Project Teams among others. Although your firm’s response states you will revise the relevant procedures, the revised procedures were not included in the response for evaluation. Your firm’s response does not include a retrospective review of related documentation to ensure compliance with the updated procedures. Documentation of corrective action, retrospective review as well as training is needed in order to adequately address this observation.

Other federal agencies may take your compliance with the FD&C Act and its implementation regulations into account when considering the award of federal contracts. Additionally, should FDA determine that you have Quality System regulation violations that are reasonably related to premarket approval applications for Class III devices, such devices will not be approved until the violations have been addressed.

Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, including an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective action (which must address systemic problems) that your firm has taken. If your firm’s planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Please provide a translation of documentation not in English to facilitate our review. We will notify you regarding the adequacy of your firm’s responses and the need to re-inspect your firm’s facility to verify that the appropriate corrections and/or corrective actions have been made. If you believe that your products are not in violation of the FD&C Act, include your reasoning and any supporting information for our consideration as part of your response.

Your firm’s response should be sent via email to CDRHWarningLetterResponses@fda.hhs.gov. Refer to CMS case 707156 when replying. If you have any questions about the contents of this letter, please contact: Lu Jiang at +1(240)402-5779.

Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm’s facility. It is your firm’s responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, FDA 483, issued at the close of the inspection may be symptomatic of serious problems in your firm’s manufacturing and quality management systems. Your firm should investigate and determine the causes of any violations and take prompt actions to address any violations and bring the products into compliance.

Sincerely yours,/S/

Robert Ochs, Ph.D.DirectorOHT8: Office of Radiological HealthOffice of Product Evaluation and QualityCenter for Devices and Radiological Health

CC:Devan MoserU.S. AgentSedecal USA Inc.7555 N. Caldwell AvenueNiles, Illinois 60714dmoser@sedecalusa.com

Citations and trends

Each 21 CFR section, ISO 13485 clause and issue type in this letter, how often it appears across all letters, and its trend by the year FDA issued the letters. 2026 runs to Nov 3, 2026, compared with the same dates last year. Select a row or a bar to see those letters in Letters. As cited in this historical letter, not a statement of current requirements.

Regulations

21 CFR 1020.30Diagnostic x-ray systems and their major components2letters cite it
0 letters issued this year vs 1 by the same date last year · -100%About 21 CFR 1020.30 →
020210202202023120241202502026
21 CFR 1002.30Records to be maintained by manufacturers1letters cite it
0 letters issued this year vs 1 by the same date last year · -100%About 21 CFR 1002.30 →
020210202202023020241202502026

Also cited: 21 CFR 806, 21 CFR 1000

Issue types

CAPAAI-assigned issue type101letters1finding here
22 letters issued this year vs 22 by the same date last year · +0%
Design controlsAI-assigned issue type90letters1finding here
18 letters issued this year vs 21 by the same date last year · -14%

Record details

Issuing center
Center for Devices and Radiological Health
Issuing office
Center for Devices and Radiological Health
Location
ALGETE, Madrid, Spain
Letter posted date
May 27, 2025
Letter issue date
May 5, 2025
Addressed to
President and CEO
Inspection dates
Jan 20, 2025 to Jan 24, 2025
Response window
15 business days after receipt of this letter
Read the source clause
Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, including an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again.
Record length
1,610 words · 0 recorded redactions
Published status
No published close-out
Original letter
View on FDA.gov (opens in a new tab)
Dataset last checked
Nov 11, 2026 · How this is built

A response window is not a calculated due date. Source excerpts are selected passages. Finding interpretations and categories are model-assisted. Verify the original letter before relying on a finding.

See the pattern across 1,707 lettersFind letters like this →
Keep exploring

More medical-device letters

Recent records for the same product type. Your previous filters remain available when you return.

Keep exploring

Find letters like this →
For regulated teams

Put this kind of reading to
work on your own records.

Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.

Every letter by dateHow this database is built

FDA Warning Letter Alerts

Share this view

Send the same filters, search and view to a colleague.

You choose what to post or send. Sharing a link does not change access to this site.