Devices

Owens & Minor, Inc. dba American Contract Systems, Inc.

FDA issued a warning letter to Owens & Minor, Inc. dba American Contract Systems, Inc. (Mechanicsville, VA, United States) on Nov 1, 2024, citing 21 CFR 820.80 and 21 CFR 820.30. Subject: CGMP, QSR, Medical Devices, Adulterated.

From the letter: “On at least four occasions, from January 2024 through July 2024, your firm accepted sterilization bag lots (86149183, 86471442, 86740750, 86149178) that failed your thickness requirements.”

Mechanicsville, VA, United StatesIssued Nov 1, 2024Posted Nov 12, 2024No close-out recorded

At a glance

What a reviewer notes before reading: findings, citations, the response deadline and what the letter warns of.

Violations
2 findings · 2 issue types
Response due
15 business days after receipt of this letter
FDA asks for
7 actions
Named as possible actions
Seizure, injunction, civil money penalties
Issued by
Center for Devices and Radiological Health
Published status
No close-out recorded

AI assisted findings

Model-assisted reading of each issue. Open one to see the FDA text it comes from, and check it against the full letter.

01Production & process controls · 21 CFR 820.80(b)Firm accepted sterilization bag lots that failed thickness requirements, using them in processing.
On at least four occasions, from January 2024 through July 2024, your firm accepted sterilization bag lots (86149183, 86471442, 86740750, 86149178) that failed your thickness requirements. Your Receiving Procedure, RV-01, includes minimum and maximum acceptance criteria for these bags to ensure they are free from defects that may adversely affect gas injection, sealing, or diffusion characteristics during your ethylene oxide (EO) sterilization processes. Acceptance activities for these 4 lots were being performed by ACS personnel at sister facilities in Florida. Each of these lots were deemed acceptable and the lots were used for EO processing at this Tiffin, OH facility, as well as other ACS facilities.
02Design controls · 21 CFR 820.30(a), 21 CFR 820.30(g), 21 CFR 820.30(j)Firm failed to complete design verification and validation for 25 product codes and lacked complete design history files.
Specifically, your firm has not completed design verification and validation activities for 25 different FDA product codes in use at this facility, to ensure that all components being placed in your convenience kits have been reviewed and verified to undergo your sterilization process, as required by 21 CFR 820.30(g). For example, two out of 26 components that are included in your C-Section Pack (pack number WECS27G) have not been verified as acceptable for your EO sterilization process. The “CAUTERY| ES ROCKER COATED SMOKE EVAC NS” is classified as a “medium” risk component and there was no documentation that this component was evaluated, as required by your “design verification and validation activities” that are described in your design control procedure. Your Design Control Procedure, DE-01, requires component evaluation be performed as part of your design verification and validation process, including your C-Section Pack, designated as FDA Code OHM, class 2. The procedure also requires that all these review records be included in the design history file. Also, your firm did not have complete design history files for 25 different FDA product codes being manufactured at your facility, as required by 21 CFR 820.30(j).

Additional FDA passages

2 extracted findings

Selected FDA passages, not the complete letter.

01

Failure to establish and maintain procedures for acceptance of incoming product to ensure that incoming product is inspected, tested, or otherwise verified as conforming to specified requirements, and that acceptance or rejection is documented, as required by 21 CFR 820.80(b). For example:

02

Failure to establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met, as required by 21 CFR 820.30(a).

What FDA requested

  • Provide a copy of CAPA 2024-0021 describing preventive actions.
  • Provide a systemic response describing how the firm will prevent recurrence of accepting nonconforming sterilization bags.
  • Provide documentation of design verification and validation activities for all 25 FDA product codes.
  • Provide a plan to complete design history files for all 25 FDA product codes.
  • Provide a plan to prioritize and implement the review of all 25 product codes.
  • Provide documentation of component evaluations for all devices.
  • Provide a plan to ensure only devices with complete DHFs are distributed.

FDA Warning Letter

27 paragraphs · 1 section

Text as published on FDA.gov, retrieved Sep 8, 2026. Layout added for reading; the wording is FDA’s. View on FDA.gov (opens in a new tab)

1WARNING LETTER

23 paragraphs

CMS # 695683 Dear Mr. Pesicka:

During an inspection of your firm, American Contract Systems, (ACS) Inc., in Tiffin OH, FEI 3008492337, on August 5 through 14, 2024, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures sterile convenience kits, such as the C-Section Pack (pack number WECS27G). Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or any function of the body.

This inspection revealed that these devices are adulterated within the meaning of section 501(h) of the Act, 21 U.S.C. § 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the current good manufacturing practice requirements of the Quality System regulation found at Title 21, Code of Federal Regulations (CFR), Part 820.

We received a response from Daniel R. Wright, VP, Global QARA, Owens & Minor, Inc., dated September 5, 2024, concerning our investigator’s observations noted on the Form FDA 483 (FDA 483), List of Inspectional Observations, that was issued to your firm. We address this response below, in relation to each of the noted violations. These violations include, but are not limited to, the following:

1. Failure to establish and maintain procedures for acceptance of incoming product to ensure that incoming product is inspected, tested, or otherwise verified as conforming to specified requirements, and that acceptance or rejection is documented, as required by 21 CFR 820.80(b). For example:

On at least four occasions, from January 2024 through July 2024, your firm accepted sterilization bag lots (86149183, 86471442, 86740750, 86149178) that failed your thickness requirements. Your Receiving Procedure, RV-01, includes minimum and maximum acceptance criteria for these bags to ensure they are free from defects that may adversely affect gas injection, sealing, or diffusion characteristics during your ethylene oxide (EO) sterilization processes. Acceptance activities for these 4 lots were being performed by ACS personnel at sister facilities in Florida. Each of these lots were deemed acceptable and the lots were used for EO processing at this Tiffin, OH facility, as well as other ACS facilities.

We reviewed your firm’s response and concluded that it is not adequate. You indicated that CAPA 2024-0021 was opened but did not provide a copy that describes the preventive actions you are planning to address this serious observation.

Your response states that on August 6, 2024, you placed a hold on all raw material sterilization bags at all ACS locations, including Tiffin, OH. All finished goods in ACS manufacturing locations and O&M distribution centers were then placed on distribution hold on August 7, 2024, while you investigated these failures. Your investigation included performing a retest of all available bags in inventory at your Fort Myers facility and you indicated that all retests passed your specifications. However, your response did not provide a systemic response describing how your firm will prevent this significant failure from recurring.

(b)(4)

2. Failure to establish and maintain procedures to control the design of the device in order to ensure that specified design requirements are met, as required by 21 CFR 820.30(a).

You did not implement your Design Control Procedure, DE-01.

Specifically, your firm has not completed design verification and validation activities for 25 different FDA product codes in use at this facility, to ensure that all components being placed in your convenience kits have been reviewed and verified to undergo your sterilization process, as required by 21 CFR 820.30(g).

For example, two out of 26 components that are included in your C-Section Pack (pack number WECS27G) have not been verified as acceptable for your EO sterilization process. The “CAUTERY| ES ROCKER COATED SMOKE EVAC NS” is classified as a “medium” risk component and there was no documentation that this component was evaluated, as required by your “design verification and validation activities” that are described in your design control procedure.

Your Design Control Procedure, DE-01, requires component evaluation be performed as part of your design verification and validation process, including your C-Section Pack, designated as FDA Code OHM, class 2. The procedure also requires that all these review records be included in the design history file.

Also, your firm did not have complete design history files for 25 different FDA product codes being manufactured at your facility, as required by 21 CFR 820.30(j).

We reviewed your firm’s response and concluded that it is not adequate. You acknowledged documentation indicating component approval for products was lacking at this facility and were not part of your DHF’s. You now plan to remediate all components and approve all DHFs, by January 31, 2025, in part to ensure that all components being placed in your convenience kits have been reviewed and verified to undergo your sterilization process. Your firm did not provide enough detail to describe how you plan on prioritizing and implementing this review for all 25 product codes in use at this location.

(b)(4). As part of your response to this observation, please describe how you will complete this review of components to ensure that only devices with complete DHF’s, including complete component evaluations, are currently being distributed.

(b)(4)

Your firm should take prompt action to address any violations identified in this letter. Failure to adequately address this matter may result in regulatory action being initiated by the FDA without further notice. These actions include, but are not limited to, seizure, injunction, and civil money penalties.

Other federal agencies may take your compliance with the FD&C Act and its implementing regulations into account when considering the award of federal contracts. Additionally, should FDA determine that you have Quality System regulation violations that are reasonably related to premarket approval applications for Class III devices, such devices will not be approved until the violations have been addressed. Should FDA determine that your devices or facilities do not meet the requirements of the Act, requests for Certificates to Foreign Governments (CFG) may not be granted.

Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, as well as an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective actions (which must address systemic problems) that your firm has taken. If your firm’s planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Your firm’s response should be comprehensive and address any violations included in this Warning Letter. If you believe that your products are not in violation of the FD&C Act, include your reasoning and any supporting information for our consideration as part of your response.

Your firm’s response should be sent by email to CDRHWarningLetterResponses@fda.hhs.gov. Refer to CMS case # 695683 when replying. If you have any questions about the contents of this letter, please contact: Gina Brackett, EAT Assistant Director at (513) 288-7055 or gina.brackett@fda.hhs.gov.

Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm’s facility. It is your firm’s responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, FDA 483, issued at the close of the inspection may be symptomatic of serious problems in your firm’s manufacturing and quality management systems. Your firm should investigate and determine the causes of any violations and take prompt actions to address any violations and bring the products into compliance.

Sincerely yours, /S/

RDML Sean M. Boyd, MPH, USPHS Director Office of Regulatory Programs Office of Product Evaluation and Quality Center for Devices and Radiological Health

Cc: (b)(4) , Operations Manager, (b)(4) @owens-minor.com

Citations and trends

Each 21 CFR section, ISO 13485 clause and issue type in this letter, how often it appears across all letters, and its trend by the year FDA issued the letters. 2026 runs to Nov 3, 2026, compared with the same dates last year. Select a row or a bar to see those letters in Letters. As cited in this historical letter, not a statement of current requirements.

Regulations

Issue types

Design controlsAI-assigned issue type90letters1finding here
18 letters issued this year vs 21 by the same date last year · -14%

Record details

Issuing center
Center for Devices and Radiological Health
Issuing office
Center for Devices and Radiological Health
Location
Mechanicsville, VA, United States
Letter posted date
Nov 12, 2024
Letter issue date
Nov 1, 2024
Addressed to
President & Chief Executive Officer
Inspection dates
Aug 5, 2024
Response window
15 business days after receipt of this letter
Read the source clause
Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, as well as an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again.
Record length
1,360 words · 5 recorded redactions
Published status
No published close-out
FDA reference
695683
Original letter
View on FDA.gov (opens in a new tab)
Dataset last checked
Nov 11, 2026 · How this is built

A response window is not a calculated due date. Source excerpts are selected passages. Finding interpretations and categories are model-assisted. Verify the original letter before relying on a finding.

See the pattern across 1,707 lettersFind letters like this →
Keep exploring

More medical-device letters

Recent records for the same product type. Your previous filters remain available when you return.

Keep exploring

Find letters like this →
For regulated teams

Put this kind of reading to
work on your own records.