Ming Fai Industrial Co LTD

FDA issued a warning letter to Ming Fai Industrial Co LTD (Shenzhen Shi, Guangdong Sheng, China) on Jun 1, 2023, citing 21 CFR 820.75 and 21 CFR 820.86. Subject: CGMP, QSR, Medical Devices, Adulterated.

From the letter: “Failure to identify by suitable means the acceptance status of the product, to indicate the conformance or nonconformance of product with acceptance criteria, as required by 21 CFR 820.86.”

Shenzhen Shi, Guangdong Sheng, ChinaIssued Jun 1, 2023Posted Nov 14, 2023No close-out recorded

At a glance

What a reviewer notes before reading: findings, citations, the response deadline and what the letter warns of.

Violations
2 findings · 1 issue type
Response due
15 business days after receipt of this letter
FDA asks for
6 actions
Named as possible actions
None named
Issued by
Center for Devices and Radiological Health
Published status
No close-out recorded

AI assisted findings

Model-assisted reading of each issue. Open one to see the FDA text it comes from, and check it against the full letter.

01Production & process controls · 21 CFR 820.75(a)Process validation documentation was missing for EO sterilization, formulation software, and cleaning of mixing tank equipment.
Failure to adequately ensure that when the results of a process cannot be fully verified by subsequent inspection and test, the process shall be validated with a high degree of assurance and approved according to established procedure, as required by 21 CFR 820.75(a). For example, your firm did not maintain documentation of validation for the following processes: a. Ethylene Oxide (EO) sterilization of the probe cover kit containing the sterile gel pouch - Your firm’s EO sterilization validation was conducted for a kit containing another device, C-Arm Drape, that is also packaged in a Tyvek pouch. However, the C-Arm Drape kit differs from the ultrasound gel probe cover kit, as the probe cover kit contains a sterile gel pouch. Your firm did not maintain documentation with justification that the EO validation conducted on the C-Arm Drapes packaged in a Tyvek pouch is representative for the probe cover kit. b. Formulation calculation software for mixing of the ultrasound gel - Your firm did not maintain validation records for the software used to calculate the amounts for the (b)(4) ingredients to be mixed for the ultrasound gel formulation. c. Cleaning of the mixing tank equipment used for ultrasound gel manufacture - Your firm did not maintain documentation of cleaning validation conducted for the mixing tank equipment used for manufacturing ultrasound gel.
02Production & process controls · 21 CFR 820.86Acceptance status of in-process and finished products was not adequately identified, with pallets mixed and sterility status not marked.
Failure to identify by suitable means the acceptance status of the product, to indicate the conformance or nonconformance of product with acceptance criteria, as required by 21 CFR 820.86. For example, your firm had several pallets of in-process product held next to finished products in an area designated for finished sterile product ready for release. Additionally, the sterility status of product was not identified (e.g., stickers affixed directly to finished product). Only a single loose paper with the lot number and one sterile color identifier ((b)(4)) was taped to the pallet.

What FDA requested

  • Provide documentation of the investigation and evidence showing the mixing tank was incorrectly listed on the report information.
  • Provide documentation of an updated procedure instructing the responsible QA engineer to review and check mixer code for each validation report.
  • Provide documentation of training on the new procedure.
  • Provide evidence of labeling on the rack.
  • Provide documentation demonstrating personnel were trained on the updated layout and procedure.
  • Provide documentation of a review or evaluation on the effectiveness of the layout change and revised Finished Device Acceptance procedure.

FDA Warning Letter

29 paragraphs · 1 section

Text as published on FDA.gov, retrieved Sep 8, 2026. Layout added for reading; the wording is FDA’s. View on FDA.gov (opens in a new tab)

DatedJune 1, 2023
AddressedDear Mr. Yim

1Letter text

23 paragraphs

During an inspection of your firm, located in Shenzhen, China, on March 06, 2023, through March 10, 2023, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures ultrasound gel and probe cover products. Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or function of the body.

This inspection revealed that these devices are adulterated within the meaning of section 501(h) of the Act, 21 U.S.C. § 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the current good manufacturing practice requirements of the Quality System regulation found at Title 21, Code of Federal Regulations (CFR), Part 820.

We received a response dated March 23, 2023, concerning our investigator’s observations noted on the Form FDA 483 (FDA 483), List of Inspectional Observations, which was issued to your firm. We address this response below, in relation to each of the noted violations. These violations include, but are not limited to, the following:

1. Failure to adequately ensure that when the results of a process cannot be fully verified by subsequent inspection and test, the process shall be validated with a high degree of assurance and approved according to established procedure, as required by 21 CFR 820.75(a). For example, your firm did not maintain documentation of validation for the following processes:

a. Ethylene Oxide (EO) sterilization of the probe cover kit containing the sterile gel pouch

Your firm’s EO sterilization validation was conducted for a kit containing another device, C-Arm Drape, that is also packaged in a Tyvek pouch.

However, the C-Arm Drape kit differs from the ultrasound gel probe cover kit, as the probe cover kit contains a sterile gel pouch. Your firm did not maintain documentation with justification that the EO validation conducted on the C-Arm Drapes packaged in a Tyvek pouch is representative for the probe cover kit.

b. Formulation calculation software for mixing of the ultrasound gel

Your firm did not maintain validation records for the software used to calculate the amounts for the (b)(4) ingredients to be mixed for the ultrasound gel formulation.

c. Cleaning of the mixing tank equipment used for ultrasound gel manufacture

Your firm did not maintain documentation of cleaning validation conducted for the mixing tank equipment used for manufacturing ultrasound gel.

We reviewed your firm’s response and concluded that it is not adequate. Your firm provided four updates corresponding to the listed examples regarding process validation.

a. Your firm provided a new procedure, (b)(4) Procedure for Irradiation Sterilization Dose, to document the irradiation validation dose similar to the validation report in Exhibit 22. However, this document does not address the concern of your firm not being able to provide adequate records of the gel lots used during the sterilization validation. Additionally, there is no evidence of procedure effectiveness and documentation of training on the new procedure.

b. Your firm provided a procedure for “Evaluate the Medical Device Family for EO sterilization” under procedure (b)(4). The purpose of this procedure is for your firm to evaluate the sterilization methods and parameters of devices to determine how to select a representative model for sterilization of a device family. However, the evaluation has not been completed by your firm and only the procedure to evaluate the representative device has been provided. The response is inadequate because your firm has not provided documentation on training and effectiveness of the new procedure. Therefore, the Agency cannot determine if your firm’s corrections adequately address the requirements of the regulation.

c. Your firm established a procedure “Software Validation for formulation” (b)(4) for the “validation of formulation systems.” The procedure documents how an Installation Qualification (IQ), Operational Qualification (OQ), and Performance Qualification (PQ) should be performed. However, the response is inadequate because of the following: - There is no specific documentation of the IQ/OQ/PQ protocol and report related to the formulation of the ultrasound gel that include acceptance criteria. - No CAPA to track and evaluate effectiveness of corrective actions - No documentation of training on the new procedure d. Your firm states that their cleaning validation incorrectly did not list the mixing tank used in the manufacture of ultrasound gel. In your response, your firm provided an equipment list that only lists the mixing tank used for the ultrasound gel manufacturing process. Your firm states that the assigned QA engineer is responsible to review and check the mixer code for each validation report. However, the response is inadequate because of the following: - There is no documentation of the investigation and evidence to show that the mixing tank was incorrectly listed on the report information. - No documentation of any updated procedure that instructs the responsible QA engineer to review and check mixer code for each validation report. - No documentation of training on new procedure.

2. Failure to identify by suitable means the acceptance status of the product, to indicate the conformance or nonconformance of product with acceptance criteria, as required by 21 CFR 820.86. For example, your firm had several pallets of in-process product held next to finished products in an area designated for finished sterile product ready for release. Additionally, the sterility status of product was not identified (e.g., stickers affixed directly to finished product). Only a single loose paper with the lot number and one sterile color identifier ((b)(4)) was taped to the pallet.

We reviewed your firm’s response and concluded that it is not adequate. Your firm provided an updated warehouse layout where you proposed storage of sterilized semi-finished products and stated that “labeling on the rack is corrected immediately”. Additionally, your firm has provided a revised procedure for “Finished Device Acceptance, (b)(4)”. However, the response is inadequate because of the following:

a. No evidence of the labeling on the rack was provided. b. No documentation was provided demonstrating personnel were trained on the updated layout and procedure. c. No documentation of a review or evaluation on the effectiveness of the layout change and revised Finished Device Acceptance procedure.

Other federal agencies may take your compliance with the FD&C Act and its implementation regulations into account when considering the award of federal contracts. Additionally, should FDA determine that you have Quality System regulation violations that are reasonably related to premarket approval applications for Class III devices, such devices will not be approved until the violations have been addressed.

Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, including an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective action (which must address systemic problems) that your firm has taken. If your firm’s planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Please provide a translation of documentation not in English to facilitate our review. We will notify you regarding the adequacy of your firm’s response(s) and the need to re-inspect your firm’s facility to verify that the appropriate corrections and/or corrective actions have been made. If you believe that your products are not in violation of the FD&C Act, include your reasoning and any supporting information for our consideration as part of your response.

Your firm’s response should be sent via email to CDRHWarningLetterResponses@fda.hhs.gov.

Refer to CMS case #660075 when replying. If you have any questions about the contents of this letter, please contact: Yanna Kang, Ph.D. at +1(301)796-6704.

Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm’s facility. It is your firm’s responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, FDA 483, issued at the close of the inspection may be symptomatic of serious problems in your firm’s manufacturing and quality management systems. Your firm should investigate and determine the causes of the violations, and take prompt actions to correct the violations and bring the products into compliance.

Sincerely, /S/

Robert Ochs, Ph.D. Director OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Cc: David Lennarz U.S. Agent Registrar Corp. 144 Research Drive Hampton, Virginia 23666 David.Lennarz@registrarcorp.com

Citations and trends

Each 21 CFR section, ISO 13485 clause and issue type in this letter, how often it appears across all letters, and its trend by the year FDA issued the letters. 2026 runs to Oct 30, 2026, compared with the same dates last year. Select a row or a bar to see those letters in Letters. As cited in this historical letter, not a statement of current requirements.

Regulations

21 CFR 820.86Acceptance status7letters cite it
0 letters issued this year vs 3 by the same date last year · -100%About 21 CFR 820.86 →

Issue types

Record details

Issuing center
Center for Devices and Radiological Health
Issuing office
Center for Devices and Radiological Health
Location
Shenzhen Shi, Guangdong Sheng, China
Letter posted date
Nov 14, 2023
Letter issue date
Jun 1, 2023
Addressed to
Quality Director
Inspection dates
Mar 6, 2023 to Mar 10, 2023
Response window
15 business days after receipt of this letter
Read the source clause
Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, including an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again.
Record length
1,473 words · 6 recorded redactions
Published status
No published close-out
Original letter
View on FDA.gov (opens in a new tab)
Dataset last checked
Nov 9, 2026 · How this is built

A response window is not a calculated due date. Source excerpts are selected passages. Finding interpretations and categories are model-assisted. Verify the original letter before relying on a finding.

See the pattern across 1,703 lettersFind letters like this →
Keep exploring

More medical-device letters

Recent records for the same product type. Your previous filters remain available when you return.

Keep exploring

Find letters like this →
For regulated teams

Put this kind of reading to
work on your own records.

Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.

Every letter by dateHow this database is built

FDA Warning Letter Alerts

Share this view

Send the same filters, search and view to a colleague.

You choose what to post or send. Sharing a link does not change access to this site.