Genovate
3 extracted findings · 2 model-assigned issue types
FDA source excerpt, selected passage
FDA has determined that your HIV DBS card self-collection kit is adulterated under section 501(f)(1)(B) of the FD&C Act, 21 U.S.C. 351(f)(1)(B) because it does not have an approved premarket approval application (PMA) in effect pursuant to section 515(a) of the FD&C Act, 21 U.S.C. 360e(a), or an approved application…
Findings and supporting evidence
Model-assisted readings, not an exhaustive interpretation. Compare with the extracted source findings below.
The kit is adulterated because it lacks an approved premarket approval application or an investigational device exemption.
Supporting FDA text
Source excerpt · automated text-match checkFDA has determined that your HIV DBS card self-collection kit is adulterated under section 501(f)(1)(B) of the FD&C Act, 21 U.S.C. 351(f)(1)(B) because it does not have an approved premarket approval application (PMA) in effect pursuant to section 515(a) of the FD&C Act, 21 U.S.C. 360e(a), or an approved application for an investigational device exemption (IDE) under section 520(g) of the FD&C Act, 21 U.S.C. 360j(g).
The kit is misbranded because FDA was not notified of intent to introduce it into commercial distribution as required by section 510(k).
Supporting FDA text
Source excerpt · automated text-match checkYour HIV DBS card self-collection kit is also misbranded under section 502(o) the FD&C Act, 21 U.S.C. 352(o), because notification of the intent to introduce the device into commercial distribution has not been provided to the agency, as required by section 510(k) of the FD&C Act, 21 U.S.C. 360(k).
Self-collection raises concerns about proper specimen collection, shipping, and sample stability that may affect the kit's safety and effectiveness.
Supporting FDA text
Source excerpt · automated text-match checkSelf-collection raises several issues of importance, including, for example, whether the user can safely and properly collect the specimen, if the sample has been properly shipped, and if there is adequate stability of the specimen given the time lapse between collection and testing and the potential impact of shipping conditions.
Findings extracted from the letter
Selected FDA passages, not the complete letter.
Source passage selected by the model
Source passage selected by the model
Source passage selected by the model
What FDA requested
Model-assisted summary. Consult the letter for complete requests and conditions.
- Notify FDA in writing of the specific steps taken to correct any violations.
- Provide a respon including an explanation of each step to prevent recurrence of violations and related documentation.
- (if not complete within 15 days) state reason for delay and date by which correction will be completed.