Flextronics America LLC
2 extracted findings · 0 model-assigned issue types
FDA source excerpt, selected passage
Failure to adequately ensure that when the results of a process cannot be fully verified by subsequent inspection and test that the process shall be validated with a high degree of assurance and approved according to established procedure, as required by 21 CFR 820.75(a).
Findings and supporting evidence
Model-assisted readings, not an exhaustive interpretation. Compare with the extracted source findings below.
No model-assisted finding is available. Use the extracted findings and original source.
Findings extracted from the letter
Selected FDA passages, not the complete letter.
Source passage selected by the model
Source passage selected by the model
What FDA requested
Model-assisted summary. Consult the letter for complete requests and conditions.
- Provide a comprehensive response addressing all violations, including specific steps taken to address them and plans to prevent recurrence.
- Provide updates on the revalidation activities for the puck assembly process, including a summary with supporting documentation upon completion.
- Provide an update on the verification of effectiveness of the CAPA for the broken components or connections when available.
Sections and clauses cited
As cited in this historical letter, not a statement of current regulatory requirements.