Devices

Contec Medical Systems Co., Ltd.

FDA issued a warning letter to Contec Medical Systems Co., Ltd. (Qinhuangdao Shi, Hebei Sheng, China) on Oct 2, 2025, citing 21 CFR 820.100, 21 CFR 820.30, 21 CFR 820.198, 21 CFR 820.180 and 1 more regulation. Subject: CGMP, QSR, Medical Devices, Adulterated.

From the letter: “The cause of nonconformities identified are not adequately investigated.”

Qinhuangdao Shi, Hebei Sheng, ChinaIssued Oct 2, 2025Posted Oct 28, 2025No close-out recorded

At a glance

What a reviewer notes before reading: findings, citations, the response deadline and what the letter warns of.

Violations
13 findings · 5 issue types
Response due
15 business days after receipt of this letter
FDA asks for
7 actions
Named as possible actions
Detention, refusal of admission
Issued by
Center for Devices and Radiological Health
Published status
No close-out recorded

AI assisted findings

Model-assisted reading of each issue. Open one to see the FDA text it comes from, and check it against the full letter.

01CAPA · 21 CFR 820.100(a)The firm failed to adequately investigate the cause of nonconformities.
The cause of nonconformities identified are not adequately investigated.
02CAPA · 21 CFR 820.100(a)Corrective actions were not documented according to the firm's CAPA procedure.
Corrective actions taken are not documented per your firm’s CAPA procedure.
03CAPA · 21 CFR 820.100(a)Preventive action was not documented as completed before closing a CAPA.
The preventive action of creating the procedure, “Cybersecurity Management Plan, Version 1, Effective 6/5/2025” was not documented as completed prior to closing CAPA number 20250227M on 3/18/2025.
04CAPA · 21 CFR 820.100(a)Verification of CAPA effectiveness was not documented, including for an open recall.
Verification activities performed to verify corrective and preventive actions were effective are not adequately documented. Specifically, you have not verified that customers are correctly interpreting actions to be taken in response to your open recall (Z-1764-2025).
05Design controls · 21 CFR 820.30(g)Software design validation test plans lacked defined operating conditions and acceptance criteria.
Test plans created to validate device software designs do not ensure validation testing is performed under adequately defined operating conditions. For example, test plans do not include established acceptance criteria prior to testing.
06Design controls · 21 CFR 820.30(g)The firm failed to document a patch as a deviation from the test plan.
Two security risks were found in the B-Ultrasound Diagnostic testing and your firm failed to document the patch created to shut down the function of the ports as a deviation from the test plan.
07Design controls · 21 CFR 820.30(g)Security test report omitted required testing identified in the test plan.
The Security Test Report for the Color Doppler Ultrasound Diagnostic System did not include Network Segmentation, Zone Boundary Protection, and General Purpose, person-to-person communication restriction testing as identified as required in the Test Plan.
08Design controls · 21 CFR 820.30(g)Software design validation test plans were not documented in device DHFs.
Software design validations are not documented in device DHFs. For example, your firm did not have the software Test Plans for the B-Ultrasound Diagnostic System cleared under K232895 or the software Test Plan for the Color Doppler Ultrasound Diagnostic System cleared under K232908 in the respective DHF.
09Design controls · 21 CFR 820.30(g)The firm failed to adequately implement its risk management procedure for software.
Your firm failed to adequately implement your risk management procedure for the CMS 8000 software.
10Complaints & adverse event reporting · 21 CFR 820.198(a)Complaints from eCommerce and international officials were not reviewed or evaluated for reportability.
complaints received via eCommerce sites and complaints received via international country officials are not reviewed and evaluated for reportability. Specifically, several complaints from eCommerce sites and one from Canada’s Medical Devices and Clinical Compliance Directorate were not logged in the complaint log for handling via your firm’s complaint handling procedure and were not evaluated for reportability to the FDA.
11Complaints & adverse event reporting · 21 CFR 820.198(a)Feedback on user error or component degradation was not documented as complaints.
Additionally, feedback associated with user error or known product quality degradation associated with component longevity (cables, power cord, USB cable, etc.) are not documented as complaints for trending or investigation.
12Records & data integrity · 21 CFR 820.180Records were not maintained at a location reasonably accessible to FDA officials.
the inspector asked to be taken to the location where the software test plan for the B-Ultrasound Diagnostic System could be found and your firm's Deputy Vice Manager for Certification and Registration indicated that test plans are not maintained at your firm for validation testing performed by third party vendors.
13Labeling, claims & marketing · 21 CFR 806.10(c)(12)The firm failed to include all communications in a correction/removal report.
Failure to include a copy of all communications regarding the correction or removal in the report, as required by 21 CFR 806.10(c)(12). For example, you included a paragraph in the customer notification letter sent to consignees for an open recall (Z-1764-2025) that was omitted from the version of the letter sent to the FDA as part of the written report.

Additional FDA passages

2 extracted findings

Selected FDA passages, not the complete letter.

01

Failure to adequately establish procedures for corrective and preventive action, as required by 21 CFR 820.100(a). For example,

02

Failure to adequately establish and maintain procedures for validating the device design, as required by 21 CFR 820.30(g). For example:

What FDA requested

  • Perform retroactive review of CAPAs to ensure adequate investigations and documentation.
  • Provide evidence of verification of CAPA effectiveness for recall Z-1764-2025.
  • Submit complete test reports for design validation of software devices.
  • Conduct retroactive review of complaints from eCommerce and international officials.
  • Provide evidence of systemic implementation of revised complaint handling procedures.
  • Provide evidence of systemic implementation of revised records maintenance work instruction.
  • Include all communications regarding correction or removal in recall report.

FDA Warning Letter

37 paragraphs · 3 sections

Text as published on FDA.gov, retrieved Oct 22, 2026. Layout added for reading; the wording is FDA’s. View on FDA.gov (opens in a new tab)

DatedOctober 2, 2025
AddressedDear Kun “James” Hu

1Introduction

1 paragraph

During an inspection of your firm located in Qinhuangdao, Hebei, China on June 9, 2025 through June 12, 2025, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures CMS 8000 Patient Monitor, CMS600P2PLUS B-Ultrasound Diagnostic System, CMS1700B Color Doppler Ultrasound Diagnostic System, CMS 9000 Patient Monitor, PO30 Pulse Oximeter, Pocket Fetal Doppler. Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), these products are devices because they are intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or function of the body.

2Quality System Regulation Violations

21 paragraphs

This inspection revealed that these devices are adulterated within the meaning of section 501(h) of the Act, 21 U.S.C. § 351(h), in that the methods used in, or the facilities or controls used for, their manufacture, packing, storage, or installation are not in conformity with the current good manufacturing practice requirements of the Quality System regulation found at Title 21, Code of Federal Regulations (CFR), Part 820.

We received a response from Kun “James” Hu, CEO dated July 1, 2025, concerning our investigator’s observations noted on the Form FDA 483 (FDA 483), List of Inspectional Observations, which was issued to your firm. We address this response below, in relation to each of the noted violations. These violations include, but are not limited to, the following:

1. Failure to adequately establish procedures for corrective and preventive action, as required by 21 CFR 820.100(a). For example,

A. The cause of nonconformities identified are not adequately investigated.

B. Corrective actions taken are not documented per your firm’s CAPA procedure.

C. The preventive action of creating the procedure, “Cybersecurity Management Plan, Version 1, Effective 6/5/2025” was not documented as completed prior to closing CAPA number 20250227M on 3/18/2025.

D. Verification activities performed to verify corrective and preventive actions were effective are not adequately documented. Specifically, you have not verified that customers are correctly interpreting actions to be taken in response to your open recall (Z-1764-2025).

The response dated July 1, 2025, is not adequate. In "FEI 3006979678,OBS1,WS2025061201-English" you provided the revised procedures, training records, and reopened the cited CAPAs for further corrective actions. However, you have not performed a retroactive review of your CAPAs to ensure you have adequately investigated the cause of nonconformities, documented corrective actions taken per your updated CAPA procedure, and documented the verification activities performed to verify corrective and preventive actions were effective. Further, your responses do not address the verification of corrective and preventive action effectiveness for your open recall (Z-1764-2025). As such, you have not provided evidence to support the adequate systemic implementation of your revised procedures and adequate CAPA procedures in accordance with 21 CFR 820.100(a).

2. Failure to adequately establish and maintain procedures for validating the device design, as required by 21 CFR 820.30(g). For example:

A. Test plans created to validate device software designs do not ensure validation testing is performed under adequately defined operating conditions. For example, test plans do not include established acceptance criteria prior to testing.

B. Test reports do not confirm that design outputs meet design inputs. For example,

i. Two security risks were found in the B-Ultrasound Diagnostic testing and your firm failed to document the patch created to shut down the function of the ports as a deviation from the test plan.

ii. The Security Test Report for the Color Doppler Ultrasound Diagnostic System did not include Network Segmentation, Zone Boundary Protection, and General Purpose, person-to-person communication restriction testing as identified as required in the Test Plan.

C. Software design validations are not documented in device DHFs. For example, your firm did not have the software Test Plans for the B-Ultrasound Diagnostic System cleared under K232895 or the software Test Plan for the Color Doppler Ultrasound Diagnostic System cleared under K232908 in the respective DHF.

D. Your firm failed to adequately implement your risk management procedure for the CMS 8000 software.

The adequacy of the response dated July 1, 2025, cannot be determined at this time.

While your response provided updated procedures and test plans, it appears your response did not include complete test reports. Therefore, you have not provided evidence to support the adequate systemic implementation of your revised procedures and adequate design validation in accordance with 21 CFR 820.30(g).

3. Failure to adequately establish procedures for receiving, reviewing, and evaluating complaints by a formally designated unit, as required by 21 CFR 820.198(a). For example, complaints received via eCommerce sites and complaints received via international country officials are not reviewed and evaluated for reportability. Specifically, several complaints from eCommerce sites and one from Canada’s Medical Devices and Clinical Compliance Directorate were not logged in the complaint log for handling via your firm’s complaint handling procedure and were not evaluated for reportability to the FDA. Additionally, feedback associated with user error or known product quality degradation associated with component longevity (cables, power cord, USB cable, etc.) are not documented as complaints for trending or investigation.

The response dated July 1, 2025, is not adequate. Your response included updated procedures and training records; however, you have not performed a retroactive review of complaints received through eCommerce sites or international country officials to ensure you have adequately received, reviewed, and evaluated these complaints. Further, you have not provided evidence to support the adequate systemic implementation of your revised procedures and adequate complaint receiving, reviewing, and evaluating in accordance with 21 CFR 820.198(a).

4. Failure to adequately maintain records at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA, as required by 21 CFR 820.180. For example: the inspector asked to be taken to the location where the software test plan for the B-Ultrasound Diagnostic System could be found and your firm's Deputy Vice Manager for Certification and Registration indicated that test plans are not maintained at your firm for validation testing performed by third party vendors.

The adequacy of your response dated July 1, 2025, cannot be determined at this time. Your response revised the Product Certification and Registration Work Instruction and provided the training records on the revised instruction. However, you have not provided evidence to support the adequate systemic implementation of your revised work instruction and adequate records maintenance in accordance with 21 CFR 820.180.

3Corrections and Removals Violation

8 paragraphs

Our inspection also revealed that your firm’s CMS8000 Patient Monitor devices are misbranded under section 502(t)(2) of the Act, 21 U.S.C. § 352(t)(2), in that your firm failed or refused to furnish material or information respecting the device that is required by or under section 519 of the Act, 21 U.S.C. § 360i, and 21 CFR Part 806 – Medical Devices; Reports of Corrections and Removals. Significant violations include, but are not limited to, the following:

Failure to include a copy of all communications regarding the correction or removal in the report, as required by 21 CFR 806.10(c)(12). For example, you included a paragraph in the customer notification letter sent to consignees for an open recall (Z-1764-2025) that was omitted from the version of the letter sent to the FDA as part of the written report.

Given the serious nature of the violations of the Act, all devices manufactured by your firm are subject to refusal of admission under section 801(a) of the Act, 21 U.S.C. § 381(a), in that they appear to be adulterated. As a result, FDA is taking steps to refuse entry of these devices into the United States, known as “detention without physical examination,” until these violations are addressed. In order to remove the devices from detention, your firm should provide a written response to this Warning Letter as described below and address the violation(s) described in this letter. We will notify you regarding the adequacy of your firm’s response(s) and the need to re-inspect your firm’s facility to verify that the appropriate corrections and/or corrective actions have been made.

Other federal agencies may take your compliance with the FD&C Act and its implementing regulations into account when considering the award of federal contracts. Additionally, should FDA determine that you have Quality System regulation violations that are reasonably related to premarket approval applications for Class III devices, such devices will not be approved until the violations have been addressed.

Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, including an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective actions (which must address systemic problems) that your firm has taken. If your firm’s planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Please provide a translation of documentation not in English to facilitate our review. If you believe that your products are not in violation of the FD&C Act, include your reasoning and any supporting information for our consideration as part of your response.

Your firm’s response should be sent via email to Jennifer Kozen at CDRHEnforcement@fda.hhs.gov. Please include in the subject line, “CMS Case 675091 when replying. If you have any questions about the contents of this letter, please contact: Jennifer Kozen at 301-796-5813 or Jennifer.Kozen@fda.hhs.gov.

Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm’s facility. It is your firm’s responsibility to ensure compliance with applicable laws and regulations administered by FDA. The specific violations noted in this letter and in the Inspectional Observations, FDA 483, issued at the close of the inspection may be symptomatic of serious problems in your firm’s manufacturing and quality management systems.

Your firm should investigate and determine the causes of any violations and take prompt actions to address any violations and bring the products into compliance.

Sincerely yours,/S/Bram Zuckerman Director Office of Cardiovascular DevicesOffice of Product Evaluation and Quality Center for Devices and Radiological Health

CC:Marco MuMID-LINK INTERNATIONAL CO., Ltd2219 Rimlang Drive, Suite 301Bellingham, Washington, US 98226Phone: 702-209-5185Fax: 360-925-3199Email: us.agent@mid-link.net

Citations and trends

Each 21 CFR section, ISO 13485 clause and issue type in this letter, how often it appears across all letters, and its trend by the year FDA issued the letters. 2026 runs to Nov 3, 2026, compared with the same dates last year. Select a row or a bar to see those letters in Letters. As cited in this historical letter, not a statement of current requirements.

Regulations

21 CFR 820.180General requirements3letters cite it
0 letters issued this year vs 2 by the same date last year · -100%About 21 CFR 820.180 →
020211202202023020242202502026

Also cited: 21 CFR 806

Issue types

CAPAAI-assigned issue type101letters4findings here
22 letters issued this year vs 22 by the same date last year · +0%
Design controlsAI-assigned issue type90letters5findings here
18 letters issued this year vs 21 by the same date last year · -14%

Record details

Issuing center
Center for Devices and Radiological Health
Issuing office
Center for Devices and Radiological Health
Location
Qinhuangdao Shi, Hebei Sheng, China
Letter posted date
Oct 28, 2025
Letter issue date
Oct 2, 2025
Addressed to
CEO
Inspection dates
Jun 9, 2025 to Jun 12, 2025
Response window
15 business days after receipt of this letter
Read the source clause
Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, including an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again.
Record length
1,731 words · 0 recorded redactions
Published status
No published close-out
FDA reference
8000
Original letter
View on FDA.gov (opens in a new tab)
Dataset last checked
Nov 11, 2026 · How this is built

A response window is not a calculated due date. Source excerpts are selected passages. Finding interpretations and categories are model-assisted. Verify the original letter before relying on a finding.

See the pattern across 1,707 lettersFind letters like this →
Keep exploring

More medical-device letters

Recent records for the same product type. Your previous filters remain available when you return.

Keep exploring

Find letters like this →
For regulated teams

Put this kind of reading to
work on your own records.

Botable builds and runs a custom AI layer around the software you already use, so your team can do the job you hired them for.

Every letter by dateHow this database is built

FDA Warning Letter Alerts

Share this view

Send the same filters, search and view to a colleague.

You choose what to post or send. Sharing a link does not change access to this site.