CareFusion 213, LLC
4 extracted findings · 2 model-assigned issue types
FDA source excerpt, selected passage
Your firm failed to thoroughly investigate any unexplained discrepancy or failure of a batch or any of its components to meet any of its specifications, whether or not the batch has already been distributed (21 CFR 211.192).
Findings and supporting evidence
Model-assisted readings, not an exhaustive interpretation. Compare with the extracted source findings below.
The firm received over 2,500 customer complaints about products with foreign matter, missing components, and empty units, but investigations were inadequate, failing to identify root causes, extend to other batches, or conduct appropriate market actions.
Supporting FDA text
Source excerpt · automated text-match checkYour firm failed to thoroughly investigate any unexplained discrepancy or failure of a batch or any of its components to meet any of its specifications, whether or not the batch has already been distributed (21 CFR 211.192).
The firm lacked quantitative verification to determine whether 'no growth' results in sterility tests were due to an effective sterilization cycle or insufficient viable spore population in biological indicators.
Supporting FDA text
Source excerpt · automated text-match checkYour firm failed to establish laboratory controls that include scientifically sound and appropriate specifications, standards, sampling plans, and test procedures designed to assure that components, drug product containers, closures, in-process materials, labeling, and drug products conform to appropriate standards of identity, strength, quality, and purity (21 CFR 211.160(b)).
Findings extracted from the letter
Selected FDA passages, not the complete letter.
Source passage selected by the model
Source passage selected by the model
Source passage selected by the model
Source passage selected by the model
What FDA requested
Model-assisted summary. Consult the letter for complete requests and conditions.
- Provide a report that evaluates whether the program includes effective root cause analysis, ensures CAPA effectiveness, analyzes investigations trends, improves the CAPA program when needed, implements final quality unit decisions, and is comprehensive.
- Provide a comprehensive, independent retrospective review of all complaint investigations for the last three years from the initial date of inspection.
- For all complaint investigations found by the retrospective review to be deficient, to have an inconclusive root cause, or to have no root cause identified, include a thorough review of production (e.g., batch manufacturing records).
- Provide a summary of potential manufacturing root causes for each investigation and any identified CAPA (e.g., manufacturing operation improvements).
- Identify any trends across complaints that may indicate systemic manufacturing, quality control, supplier qualification, or design issues requiring corrective action.
- Provide a comprehensive, independent assessment of your bioburden control program for non-sterile components and finished drug products.
- Provide bioburden and sterility testing data for all finished product batches manufactured over the last three years, including all data obtained when testing in-process and finished product batch units, as well as complete microbial identification.
- Provide a summary report of all bioburden and sterility data collected from the above assessments, including trend analysis, identification of any patterns or excursions, and corresponding CAPA to address any deficiencies identified.
Sections and clauses cited
As cited in this historical letter, not a statement of current regulatory requirements.