Analytical Food Laboratories, Inc.
4 extracted findings · 1 model-assigned issue type
FDA source excerpt, selected passage
Your firm failed to adequately validate the analytical methods used to test your customers’ drug products. For example, you conducted assay testing of over-the-counter (OTC) drug products, (b)(4) oral solution and (b)(4) tablets, but indicated during the inspection that your firm did not have written procedures for…
Findings and supporting evidence
Model-assisted readings, not an exhaustive interpretation. Compare with the extracted source findings below.
The firm did not validate analytical methods for testing drug products.
Supporting FDA text
Source excerpt · automated text-match checkYour firm failed to adequately validate the analytical methods used to test your customers’ drug products. For example, you conducted assay testing of over-the-counter (OTC) drug products, (b)(4) oral solution and (b)(4) tablets, but indicated during the inspection that your firm did not have written procedures for the assay method and had not validated the test methods you used.
The firm failed to perform appropriate system suitability checks before HPLC sample injections.
Supporting FDA text
Source excerpt · automated text-match checkYour high-performance liquid chromatography (HPLC) analyses for drug products lacked an appropriate system suitability determination, as outlined in USP <621>, prior to sample injections.
The firm did not adequately perform growth promotion testing of microbiological media.
Supporting FDA text
Source excerpt · automated text-match checkYour firm failed to adequately establish and follow procedures for growth promotion testing of your microbiological media to assure suitability before use. For example, growth promotion tests for media batches 06-19-23-3 and 06-19-23-5 were performed with only one indicator microorganism, Staphylococcus aureus, and did not include the other challenge microorganisms listed in USP <61>.
Findings extracted from the letter
Selected FDA passages, not the complete letter.
Source passage selected by the model
Source passage selected by the model
Source passage selected by the model
Source passage selected by the model
What FDA requested
Model-assisted summary. Consult the letter for complete requests and conditions.
- Provide comprehensive independent assessment of laboratory practices and remediation plan.
- Retrospective assessment of analytical methods for last 3 years and notify customers.
- Retrospective risk assessment of growth media for last 3 years and notify customers.
- Comprehensive assessment and remediation plan for quality unit authority and resources.
- Independent assessment of documentation practices with CAPA plan.
- Independent assessment of impact of inadequate documentation on product quality.
- Comprehensive independent assessment of change management system.
Sections and clauses cited
As cited in this historical letter, not a statement of current regulatory requirements.